Anesthesia Machine Components

On this page

Gas Supply System - Pressure Points

  • Cylinder Pressures (High):
    • O₂: 2000-2200 psi (~137 bar)
    • N₂O: 745 psi (~51 bar); constant if liquid
    • Air: 1800-2000 psi (~137 bar)
  • Pipeline Pressure (Intermediate):
    • Standard: 50-55 psi (3.4-3.8 bar)
  • Pressure Regulators (1st Stage):
    • Reduce cylinder P to ~45 psi.
    • Ensures constant, ↓P.
  • Pressure Relief Valves:
    • Prevents overpressure.
    • Cylinder: Burst disc, fusible plug.
    • Pipeline: Set > normal pipeline P. Anesthesia machine gas supply: pipeline & cylinder inlets

⭐ N₂O cylinder P (745 psi) is constant with liquid N₂O; weigh to check volume.

Flow Control & Meters - Gas Highway

  • Controls & measures gas flow to vaporizers.
  • Flow Control Valves (Needle Valves):
    • Fine-tunes O₂, N₂O, Air flow.
  • Flowmeters (Rotameters/Thorpe Tubes):
    • Tapered tubes; float shows flow. Read: bobbin top, ball center.
    • Accuracy: ±5% to ±10%.
    • 📌 O₂ flowmeter downstream (last).
    • Min. O₂ flow often 200-300 ml/min.
  • Hypoxic Guard (Link-25): Min. 25% O₂ with N₂O. Anesthesia machine flowmeter bank with oxygen last

⭐ O₂ flowmeter is last in sequence to prevent hypoxic mixture if an upstream (e.g., N₂O) flowmeter leaks.

Vaporizers - Dream Weavers

  • Function: Adds precise volatile anesthetic vapor to Fresh Gas Flow (FGF).
  • Location: Between flowmeters & Common Gas Outlet (CGO).
  • Types:
    • Variable Bypass (Plenum): FGF splits (vaporizing chamber/bypass). Temperature compensated (e.g., bimetallic strip). For Isoflurane, Sevoflurane.
    • Measured Flow (Heated): Tec6 for Desflurane (heated to 39°C, pressurized ~2 atm).
  • Safety Features: Agent-specific keyed fillers, interlock system (prevents >1 vaporizer ON), secure mounting.
  • Hazards: Tipping (liquid agent in bypass → ↑output), overfilling, wrong agent, leaks.

Anesthesia machine vaporizer with keyed filler

⭐ > Desflurane's low boiling point (22.8°C) necessitates a specially heated and pressurized vaporizer (e.g., Tec6) for consistent, predictable delivery, unlike other common volatile agents.

Breathing Circuits - Patient's Lifeline

Delivers anesthetic gases, removes CO2.

  • Types & Key Features:
    • Open: No rebreathing/reservoir (e.g., Schimmelbusch).
    • Semi-Open (Mapleson): No CO2 absorber; FGF critical.
      • Mapleson A (Magill): Best for Spontaneous Ventilation (SV). $FGF = MV$.
      • Mapleson D: Best for Controlled Ventilation (CV). $FGF = \textbf{1-2} \times MV$.
      • 📌 Mnemonic: "Dead A Live D" (D for CV, A for SV).
    • Closed/Semi-Closed (Circle System): Uses CO2 absorber (soda lime); low FGF.
      • Components: Unidirectional valves, absorber, APL valve.

⭐ Soda lime, the CO2 absorbent in circle systems, changes color (e.g., to violet) when exhausted.

Mapleson A and D Breathing Circuits

Safety Mechanisms - Vigilant Protectors

  • Oxygen Supply
    • Oxygen failure alarm (audible/visual). Low O₂ pressure alert.
    • Fail-safe valve: ↓/cuts N₂O if O₂ pressure < 25-30 psi.
    • Hypoxic guard (e.g., Link-25): minimum ~25% O₂ delivery with N₂O.
    • O₂ flush: high flow O₂ (35-75 L/min) to common gas outlet.
  • Vaporizers
    • Interlock system: one volatile agent active at a time.
    • Keyed-fillers: agent-specific, prevent misfilling.
  • Breathing Circuit
    • Pressure alarms: high (occlusion), low (disconnect).
    • APL valve: limits pressure in manual/spontaneous ventilation.
  • Gas Connections
    • PISS (cylinders), DISS (pipelines): non-interchangeable.
    • Color coding for gas identification.

⭐ Capnography: mandatory for confirming ETT placement & monitoring ventilation.

High‑Yield Points - ⚡ Biggest Takeaways

  • Hanger Yoke & PISS: Ensures correct cylinder attachment and gas-tight seal (Bodok seal).
  • Pressure Regulators: Reduce high cylinder pressure to a constant, lower machine operating pressure (≈45 psig).
  • Flowmeters (Rotameters): Measure gas flow; hypoxic guard links N₂O and O₂ flows. Read at top of bobbin.
  • Vaporizers: Deliver calibrated volatile anesthetic concentration. Agent-specific with interlock safety.
  • Oxygen Flush Valve: Provides high-flow O₂ (35-75 L/min) directly to common gas outlet, bypassing vaporizers.

Continue reading on Oncourse

Sign up for free to access the full lesson, plus unlimited questions, flashcards, AI-powered notes, and more.

CONTINUE READING — FREE

or get the app

Rezzy — Oncourse's AI Study Mate

Have doubts about this lesson?

Ask Rezzy, your AI Study Mate, to explain anything you didn't understand

Enjoying this lesson?

Get full access to all lessons, practice questions, and more.

START FOR FREE