Limited time75% off all plans
Get the app

Standard Monitoring: ECG, BP, Pulse Oximetry

Standard Monitoring: ECG, BP, Pulse Oximetry

Standard Monitoring: ECG, BP, Pulse Oximetry

On this page

ECG Monitoring - Heart's Electric Symphony

Continuously monitors cardiac electrical activity for arrhythmias & myocardial ischemia.

  • Purpose:
    • Rate & rhythm assessment.
    • Detect myocardial ischemia (ST segment changes).
    • Identify electrolyte imbalances (e.g., K⁺).
    • Monitor pacemaker function.
  • Lead Systems:
    • 3-Lead: Standard limb leads (I, II, III). Basic rhythm monitoring.
    • 5-Lead: Adds a precordial lead (usually V5). Better ischemia detection.
      • Lead II: Best for P-wave morphology & arrhythmia diagnosis.
      • Lead V5: Sensitive for anterior/lateral ischemia (LAD/LCx territory).
  • Key Intraoperative Changes:
    • Ischemia:
      • ST Depression: >1mm (subendocardial ischemia).
      • ST Elevation: >1mm (limb), >2mm (precordial) (transmural injury).
    • Arrhythmias: Bradycardia, tachycardia, AF, VT, ectopics.
    • Hyperkalemia: Peaked T waves, wide QRS.

ECG ST segment, elevation, and depression

⭐ Lead II is optimal for rhythm monitoring and P-wave detection; V5 is superior for detecting intraoperative myocardial ischemia.

BP Monitoring - Pressure Under Watch

  • Goal: Maintain tissue perfusion; guide hemodynamics.
  • Non-Invasive BP (NIBP)
    • Method: Oscillometry.
    • Cuff size: Width 40%, bladder length 80% arm circumference.
      • Too small → falsely ↑BP; Too large → falsely ↓BP.
    • Placement: Arm at heart level.
  • Invasive BP (IBP) / Arterial Line
    • "Gold Standard": Continuous beat-to-beat monitoring.
    • Indications: Major surgery, hemodynamic instability, vasoactive drug infusions.
    • Sites: Radial (check Allen's test), femoral, brachial.
    • Transducer: Level at phlebostatic axis (4th ICS, mid-axillary line).
      • Transducer high → falsely ↓BP; Transducer low → falsely ↑BP. Phlebostatic axis and transducer leveling
    • Waveform: Dicrotic notch (aortic valve closure).
  • Calculated Pressures
    • Mean Arterial Pressure (MAP): $MAP \approx DP + 1/3(SP-DP)$. Target >65 mmHg.
    • Pulse Pressure (PP): $PP = SP - DP$. Indicates stroke volume.

⭐ In IBP, overdamping (e.g., air bubbles, clots, kinks) causes underestimated Systolic BP, overestimated Diastolic BP, but MAP is least affected.

Pulse Oximetry - Oxygen's Window

  • Measures SpO2 (peripheral O2 saturation) & pulse rate. Non-invasive, continuous.

  • Principle: Spectrophotometry (Beer-Lambert law).

    • Differentiates HbO2 & Hb by absorption of two light wavelengths:
      • Red light (660 nm): Absorbed by deoxy-Hb (📌 RED D: RED light for Deoxy Hb).
      • Infrared light (940 nm): Absorbed by oxy-Hb (📌 I O: Infrared for Oxy Hb).
  • Normal SpO2: 95-100%. Critical: < 90%.

  • Placement: Finger, toe, earlobe. Neonates: foot/palm.

  • Limitations & Inaccuracies:

    • Motion, poor perfusion (shock, vasoconstriction, hypothermia).
    • Ambient light, dark nail polish (blue, black, green).
    • IV dyes (e.g., methylene blue: transiently falsely ↓ SpO2).
    • Dyshemoglobinemias:
      • Carboxyhemoglobin (COHb): Falsely ↑ SpO2 (COHb mimics HbO2 at 660 nm).
      • Methemoglobin (MetHb): SpO2 tends towards 85% regardless of true SaO2.

⭐ In carbon monoxide poisoning, pulse oximetry is unreliable, showing falsely high SpO2 because COHb absorbs light at 660 nm similarly to oxyhemoglobin (HbO2).

High‑Yield Points - ⚡ Biggest Takeaways

  • ECG: Lead II best for arrhythmia detection; V5 for myocardial ischemia.
  • NIBP cuff: width 40% of arm circumference, bladder length 80%.
  • Pulse oximetry (SpO2): measures O2 saturation via Beer-Lambert law.
  • Carboxyhemoglobin falsely ↑ SpO2; Methemoglobin falsely ↓ SpO2 towards 85%.
  • MAP = DBP + 1/3 (SBP-DBP); critical for assessing organ perfusion.
  • Key SpO2 inaccuracies: motion, poor perfusion, IV dyes (e.g., methylene blue), dark nail polish.
  • Oscillometric NIBP directly measures MAP; SBP and DBP are calculated.

Continue reading on Oncourse

Sign up for free to access the full lesson, plus unlimited questions, flashcards, AI-powered notes, and more.

CONTINUE READING — FREE

or get the app

Rezzy — Oncourse's AI Study Mate

Have doubts about this lesson?

Ask Rezzy, your AI Study Mate, to explain anything you didn't understand

Enjoying this lesson?

Get full access to all lessons, practice questions, and more.

START FOR FREE